Remote Alarm - Sanyo MDF-U2086S Instruction Manual

Ultra low temperature freezer
Table of Contents

Advertisement

OPERATING INSTRUCTIONS

Remote alarm

1. MDF-U2086S
The terminal of the remote alarm is installed at the
rear part of the frame.
Contact capacity is 2A (DC 30V).
output: normal open, connectors are connected with
N.O. terminal
2. MDF-U3086S, U5186S
The terminal of the remote alarm is installed at rear of
the machinery cell, under the electric box.
Consult our dealer or our office for installation of the
remote alarm.
Contact capacity is 2A (DC 30V) - 0.1mA (DC 1V).
3. MDF-U4086S MDF-U5086W
The terminal of the remote alarm is installed at the rear
upper of the frame.
Contact capacity is 2A (DC 30V) - 0.1mA (DC 1V).
a) output: normal open
connectors are connected with N.O. side.
b) output: normal close
connectors are connected with N.C. side.
Remote alarm
Max.
18
N.O
I/F (RS485)
Terminal for remote alarm
N.C.
COM
N.O.
Fig. 1
Terminal for power
N.O
.
N.C.
Terminal for high
temp. alarm
Fig. 2