Chamber Temperature Setting; Remote Alarm Terminal - Sanyo MDF-594C Instruction Manual

Ultra-low temperature freezer
Hide thumbs Also See for MDF-594C:
Table of Contents

Advertisement

CHAMBER TEMPERATURE SETTING

Table 1 shows the basic operation method. Perform key operations in the sequence indicated in the
table. The example in the table is based on the assumption that the temperature is -70
Note: The unit is set at the factory that the chamber temperature is -80
Table 1. Basic operation sequence (Example: Chamber temperature -70
Description of operation
1 Turn on the power switch.
Press temperature setting key
2
when changing the set temperature
during operation.
Set the second digit to 7 with the
3
numerical value shift key.
5 Press enter key.
Note:
The temperature set mode returns to the temperature display mode automatically when 90
seconds has passed without any key operation.

REMOTE ALARM TERMINAL

The terminal of the remote alarm is located at rear of the machinery room, upper side of the electric box.
The signal is contact output. The recommended contact capacity is between 2 A (DC 30 V) .
a) Contact output: normal open connect with N.O. and COM.
b) Contact output: normal close connect with N.C. and COM.
Consult our dealer or our office for installation of the remote alarm.
Key operated
The current chamber temperature
-----
is displayed.
PV/SV
The second digits blinks.
When pressed the figure of settable
digit changes.
Set temperature is memorized and
ENT
the current chamber temperature is
displayed.
41
o
C.
o
C.
o
C)
Indication after operation
14