Test Strip Specifications - Advocate Redi-Code+ BMB-EA001 User Manual

Hide thumbs Also See for Redi-Code+ BMB-EA001:
Table of Contents

Advertisement

No Coding Required!
Test Strip
Specifications
Advocate Redi-Code+ BMB-BA006A
Analysis Method
Physical Characteristics
Dimensions (mm)
Connector (pins)
Material
Base Material
Conductive Layer
Electrode
Chemical Components
Enzyme
Mediator
Others
Function
Test Sample
Sample Size
Test Time (sec.)
Performance
Linear Range
Precision
Accuracy
Environmental Condition
Operation (Temp. / Humidity)
Storage (Temp. / Humidity)
Stability
Opened vial (days)
Closed vial (months)
Warning:
- Do not use expired strip
- This strip is only used with Advocate Redi Code+ Meters
- Close the can immediately after strip has been removed
- Fluoride: Cannot put fluoride into blood as the preservative
- Iodoacetic Acid: It is impossible to put Iodoacetic Acid into blood as the preservative as well.
- New Born Baby: New born babies are not recommended to do this test.
- Reductant: High density of Vitamin C or uric acid may influence glucose value
- Hematocrit: Hematocrit above 60% may cause lower glucose results. Hematocrit lower than 20% may cause higher
glucose results.
Blood Glucose Test Strips
Electrochemical
30.9 x 6 0.58
6
PET
Ag & Carbon
Carbon
Glucose Oxidase
Electron Shuttle
Non-reactive ingredients
Whole Blood
1 µL
5 sec. / 7 sec.
20-600 mg/dL (1.1~33.3 mmol/L)
CV < 5%
<75 mg/dL ± 15mg/dL ; ≥ 75mg/dL ± 20%
10°C - 40°C / RH below 85%
4°C - 40°C / RH below 85%
90 days
24 months
5