Control Solution Test
Step 5:
Compare the result with
the range printed on the test strip
vial. The result should fall within
this range. Remove the used strip,
the meter will automatically turn off.
• The control solution range printed on the test strip vial is for Advocate
Redi-Code+ Control Solution only. It is not a recommended range for
your blood glucose test results.
• Everytime you perform a control solution test, you must enter into the
"CTL" mode to avoid storing the result in the meter memory. Failure
to do so will confuse the blood glucose result with the control solution
result in meter's memory.
Warnings:
To reduce the chance of infection
• Make sure to wash and dry the puncture site before starting.
• Never share a lancet or the lancing device with anyone.
• Always use a new, sterile lancet. Lancets are for single use only.
• Keep the meter, the lancets and the lancing device clean before using.
• The user should wash hands thoroughly with soap and water after
handling the meter, lancing device, or test strips.
• If the meter is being operated by a second person who is providing
testing assistance to the user, the meter and lancing device should be
decontaminated prior to use by the second person.