Advocate Redi-Code+ BMB-EA001 User Manual page 26

Hide thumbs Also See for Redi-Code+ BMB-EA001:
Table of Contents

Advertisement

Perform
Blood Glucose Test
Step 4: Apply Blood to the Test Strip
Apply the blood sample to the test strip tip. The
minimum blood sample size is 1.1μl (●). Blood
will be drawn into the strip. Keep holding the
drop of blood to the hole of the test strip until the
confirmation window is full and the meter begins
count down. Please note that if the
confirmation window is not filled with blood,
the test result may be inaccurate.
Step 5: Get Your Results
After 5 seconds, your blood glucose test result
appears. This result will be automatically stored
in the meter's memory.
Step 6: Dispose the Used Lancet
Remove the lancet carefully to prevent injury.
Take the lancet out from the lancing device. Place the disk
on a hard surface and push the
cover.
Warning:
- The user should wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips.
- If the meter is being operated by a second person who is providing testing assistance to the user, the meter and lancing device should
be decontaminated prior to use by the second person. Please refer to the section "Cleaning and Disinfection" for details in this manual.
- If no action is done for 2 minutes, the meter will automatically turn off.
procedure.
- The blood should completely fill the confirmation window before the meter begins to count down. If you find that the confirmation
window is not filled with blood when the meter is counting, NEVER try to add more blood to the test strip.
Discard the test strip and retest with a new one.
- Any trouble filling the confirmation window, contact your Customer Support at 1-866-373-2824 for help.
- The used lancet and the used test strip may be potential biohazard. Please discard it carefully according to your healthcare provider's
instructions.
exposed tip into the disk
You must remove and reinsert the test strip to restart the test
26

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Redi-code+ bmb-ea001s

Table of Contents