11.0 RS-232 INTERFACE

The balances have the ability to send data to the serial interface.
The weighing data can be sent over the interface either automatically or when
the user presses the [Print] key.
The user has control over what data is to be printed.
The following gives a description of the RS-232 interface.
HARDWARE
The RS-232 interface is a simple 3 wire connection. The input and output
connections are:
Connector:
Baud rate:
Parity:
All lines are terminated with carriage return and line feed (<CR><LF>).
In continuous output mode, or if single-line output on demand is selected, the
serial output format will be a single line in the form "1234.567 g<CR><LF>".
The format of the single-line output will change depending on the mode in
which the balance is operating, as described below.
If output on demand is selected, the user may optionally configure the serial
output as a choice of 3 styles of form, either in a default format or in one of
two custom formats. Each of the custom formats can be configured to output
up to 15 lines of data. The data types that can be printed are:
NAME
ID number
Serial number
Date
Time
Net weight
Gross weight
9 pin D-sub miniature socket
Pin 2 input to balance RXD
Pin 3 output from balance TXD
Pin 5 Signal ground GND
Handshaking is not applied.
4800, 9600, 19200, 38400
NONE (=8N1), EVEN (=7E1) or ODD (=7 O 1)
TEXT PRINTED
ID no.: xxxxxxxxxxxx
Serial no. xxxxxxxxxxxx
DATE dd/mm/yyyy
TIME hh:mm:ss
Net: xxx.xxx g
Gross: xxx.xxx g