Evaluating the control test result
The permissible range for the control solution
reading is indicated on the label of the test strip
container. Your test result must lie within the
range indicated. Make sure that you compare the
test result with the correct range. If the control
test result lies within the range indicated on the
test strip container, then the monitor and the
test strips are working accurately. If the control
test result does not lie within the range indicated
on the test strip container, then the following
options will be displayed to rectify the problem:
Cause
Has the test strip
been lying around
open for a long time?
Was the test strip
container properly
closed?
Remedy
Repeat the test with a
test strip that has
been stored correctly.
The test strips are damp.
Replace the test strip.
Cause
Did the monitor work
properly?
Is the control
solution soiled or
has it exceeded the
expiry date?
Were the test strips
and control solution
stored in a cool, dry
place?
Did you follow the
steps of the test
procedure properly?
5 Operating
Remedy
Repeat the test as describ-
ed in section 5.1. If pro-
blems persist, get in touch
with the service centre.
Use new control solution
to check the performance
of the monitor.
Repeat the control test
using strips and solution
that have been stored
correctly.
Repeat the test as de-
scribed in section 5.1.
If problems persist, get in
touch with the service
centre.