be selected if it is required to block the static background field and provide only an ac response.
The decision to use either a dc or ac response will depend upon several factors.
(a)
The need or otherwise to measure the static field.
(b)
The dynamic range of any associated measuring instrument.
If the analog signal is converted by a 16 bit A-D converter, this will have a resolution of ±1 in
±32768. This is just about sufficient to digitise a terrestrial field strength of 47µT to the full
resolution of the magnetic field sensor.
monotonicity over 16 bits.
If only the alternating component of the magnetic field is of interest the static component can be
rejected by including the high pass filter. To do this, remove the two dome head screws in the
base of the power supply unit and remove the lid. This will reveal the filters. A diagram on the
inside of the lid of the power supply unit shows how the filters can be individually linked for dc or
ac response. (See Figure 4)
As supplied, the high pass filter is linked out and the filter section has a dc response.

11.3 ADDITIONAL FILTERS

The output from the Mag-03 sensor will contain noise over a wide frequency band, including
break-through from the operating frequency of the sensors at 15.625 kHz. To acheive the lowest
output noise level the bandwidth should be limited, using an external filter with a sharp cut-off, to
the frequency band of interest.

11.4 OVERLOAD

Exposure of the magnetic field sensor to a flux between +120% and +200% or -120% and -400%
of the full scale range will cause the analog output to saturate. Beyond this the output will
become unpredictable, usually slewing to the opposite polarity. The magnetic field sensor will not
be harmed permanently due to overload but this situation should be avoided.

11.5 AC MEASUREMENT PRECAUTIONS

The measurement and roll-off characteristics of the magnetic field sensor ensure great precision
for measurements obtained within the specified bandwidth and freedom from errors at frequencies
above the -3dB point. It must be noted, however, that all fluxgate magnetic field sensors employ
some form of synchronous phase detector operating at a particular frequency. This gives rise to
modulation products in the presence of alternating flux close to the operating frequency. The
magnitude of the products will be equal to the magnitude of the detected flux (±3dB) for a
difference frequency of ±1.5kHz. The operating frequency of the Mag-03 series of magnetic
sensors is 31.25kHz.

11.6 ELECTROMAGNETIC COMPATIBILITY

The Mag-03PSU contains no high frequency electronics likely to cause emissions which could
affect other apparatus.
tantalum capacitors provide good filtering to minimise any emissions. The unit is also unlikely to
be affected by interference from other equipment in the normal operating environment. However,
The constant voltage charger and lead-acid battery combined with
Many 16 bit converters do not guarantee full