Bartington PSU1 Operation Manual

Bartington psu1 power supply unit operation manual
Hide thumbs Also See for PSU1:

Advertisement

Innovation in Magnetic Field Measuring Instruments
Operation Manual for
PSU1 Power Supply Unit
www.bartington.com

Advertisement

Table of Contents
loading

Summary of Contents for Bartington PSU1

  • Page 1 Innovation in Magnetic Field Measuring Instruments Operation Manual for PSU1 Power Supply Unit www.bartington.com...
  • Page 2: Table Of Contents

    6.1 Summary 6.2. Functional Description 6.2.1. Power 6.2.2. Signal Buffering 6.2.3. Filtering 7. PSU1 Inputs, Outputs and Controls 7.1. Back Panel Connections and Controls 7.2. Front Panel Controls 8. Installing the PSU1 8.1. Installing and Replacing the Internal Batteries 8.2. Initial Charging of the Battery Cells 8.3.
  • Page 3 BARTINGTON INSTRUMENTS 8.5. Initial Settings 8.5.1. Magnetometer Output Type Selection 9. Using the PSU1 9.1. Environmental Precautions 9.2. Switching On and Off 10. Troubleshooting 11. Care and Maintenance 11.1. Fuses 11.2. Calibration 11.3. Cleaning 12. End of Life Disposal 12.1. Waste Electrical and Electronic Equipment (WEEE) Regulations 17...
  • Page 4: Legal Notices

    Madrid Agreement & Protocol, Norway and the United States of America. 2. How to Use this Manual This manual provides the information necessary to help customers install and operate the PSU1 Power Supply Unit from Bartington Instruments.
  • Page 5: Safe Use

    6. Introduction to the PSU1 6.1 Summary The PSU1 provides a battery backed power supply of ±12V for the Mag-03 series of three-axis fluxgate magnetic field sensors. It also contains filters for the analogue outputs of the sensor. Page 5 of 18...
  • Page 6: Functional Description

    6.2. Functional Description 6.2.1. Power The PSU1 will supply ±12V DC for powering a range of magnetometers. Please refer to the datasheet for the latest list of compatible products. The power is supplied from the internal, rechargeable batteries or from the plug-in power adapter, which is also used to charge the PSU1 batteries.
  • Page 7: Psu1 Inputs, Outputs And Controls

    BARTINGTON INSTRUMENTS 7. PSU1 Inputs, Outputs and Controls 7.1. Back Panel Connections and Controls 1. X Signal Conditioned Output. 2. Y Signal Conditioned Output. 3. Z Signal Conditioned Output. The three BNC connectors carry the conditioned analogueue output voltages. The conditioned outputs are the magnetometer X, Y and Z signals after they have been modified by the low/high pass filters and, if a balanced output magnetometer is connected, converted to single-sided signals..
  • Page 8: Front Panel Controls

    BARTINGTON INSTRUMENTS 7.2. Front Panel Controls 6. Magnetometer Output Type Selector Switch Balanced (out) Unbalanced (in) Caution: This switch must be set to match your magnetometer output type to ensure correct results (see Section 8.5.1: Magnetometer output type selection) 7. DC/AC Coupling (High Pass Filter (HPF) Control)
  • Page 9: Installing The Psu1

    A. Enclosure lid retaining screw. B. Enclosure lid retaining screw. C. Earth lead fixing (do not remove) 1. Place the PSU1 on a suitable surface with the underside facing up and the two enclosure lid retaining screws nearest to you. Page 9 of 18...
  • Page 10 BARTINGTON INSTRUMENTS 2. Using a flat-bladed screwdriver of the appropriate size, fully loosen screw [A] to allow the edge of the lid to clear the enclosure end cap. 3. Fully loosen screw B and the lid of the enclosure will lift due to the internal springs.
  • Page 11 BARTINGTON INSTRUMENTS 4. Lift the lid and place aside to gain access to the battery cell compartment. 5. Insert each cell into the compartment in turn. Always fit the negative end of the cell first. Press the cell back against spring pressure until the positive terminal can be located in the holder.
  • Page 12 BARTINGTON INSTRUMENTS 6. When all five cells are in place, locate the back edge of the lid in the enclosure. 7. Tighten screws [A] and [B] in any order. 8. To remove discharged or defective cells follow the above procedure.
  • Page 13: Initial Charging Of The Battery Cells

    However the outputs may carry some charger noise. Note: If the PSU1 is powered by the mains charger with no batteries cells installed then the charging LED may flash. This is normal and can be ignored.
  • Page 14: Connecting The Equipment

    BARTINGTON INSTRUMENTS Note: The PSU1 is inherently magnetic and should, therefore, be sited as far from any magnetometers as possible. 8.4. Connecting the Equipment Caution: Connection between the magnetometer and the power supply should not be made or broken with the power supply switched on as this could cause damage to the magnetometer.
  • Page 15: Troubleshooting

    BARTINGTON INSTRUMENTS Note: For best results, after switching on the power, leave the PSU1 for 20 minutes for the internal temperature to stabilise, before performing any measurements. 10. Troubleshooting If the PSU1 exhibits unexpected behaviour, please check possible causes in the following table:...
  • Page 16: Care And Maintenance

    Return PSU1 and sensor to Bartington Instruments for repair If the solutions suggested above do not fix the fault, contact the Bartington Instruments helpdesk. 11. Care and Maintenance The PSU1 requires no routine maintenance. There are no user serviceable parts.
  • Page 17: Fuses

    BARTINGTON INSTRUMENTS 11.1. Fuses As a safety feature the PSU1 is fitted with a thermal fuse. If the PSU1 overheats, which could occur as a result of the fitment of incorrect battery cells, the charging circuit will be permanently isolated. The PSU1 must be returned to Bartington Instruments for checks and repair.
  • Page 18 T: +44 (0)1993 706565 Bartington Instruments Limited The specifications of the products described in this brochure are subject to change without prior notice. Bartington® is a registered trademark of Bartington Instruments Ltd. 5, 10 & 11 Thorney Leys Business Park F: +44 (0)1993 774813 www.bartington.com...