E-Medic Silence TP 2 User Manual page 7

All- in-one medical panel pc
Table of Contents

Advertisement

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the e-medic Silence TP 2 &
TP 1 are used exceeds the applicable RF compliance level above, the
e-medic Silence TP 2 & TP 1 should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the e-medic Silence TP 2 & TP 1.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than V/m.
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
- 6 -