  • Page 3: About The Display

    About the Display The Dome® Ex displays are available in 2, 3, and 5 megapixels and contain a TFT LCD panel. The display’s thin film transistors, in a transmissive-type display, use an integrated cold cathode fluorescent tube (CCFT) backlight system.
  • Page 4 Dome Ex display mounted on desk stand DC power adapter* Power cords DVI cable DisplayPort cable** * Some Dome Ex displays ship with a DC adapter that operates per power switch. ** For Dome E3cHB unit only. 2 | Dome Ex Display...
  • Page 5: Display Components

    Dome E2cHB, Dome E3, and Dome E5, and 6-pin port on Dome E3cHB. • Video input. Drives the data to the display. All Dome Ex displays have a standard DVI connector. The Dome E3cHB unit also includes a DisplayPort connector port.
  • Page 6: Installing The Display

    Installing the Display Turn your computer off. Leave the power cord plugged into the grounded outlet. Use the Dome Ex display with the power adapter and video cable shipped. WARNING No modification of this equipment is allowed. In locations where 240V outlets are used, connect the Dome Ex display only to a center-tapped, 240V, single-phase supply (for Canada and the United States only).
  • Page 7 HARDWARE INSTALLATION TIPS • Access to ports. For displays mounted on a desk stand, rotate the LCD panel from portrait to landscape for better access to the ports. • Threading cables. Thread the DVI cable and power cord through the stand column. Make sure the cable and cord run through the notches.
  • Page 8: Technical Information

    Technical Information The design of the Dome® Ex digital display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to the display. Read all instructions before setting up the display.
  • Page 9: Symbol Explanations

    12,000 meters (39,400 feet), maximum in unpressurized container. RELATIVE HUMIDITY. Transport and storage 5% to 90% (noncondensing). TEMPERATURE. Transport and storage -10° to 60° C for Dome E3, Dome E3cHB, and Dome E5; -20° to 60° C for Dome E2cHB. WARNING. Technical Information | 7...
  • Page 10: Intended Use

    Dome Ex line, model Dome E5 is intended for use in the displaying and viewing of medical images, including radiographs of the breast, for review and analysis by trained medical practitioners.
  • Page 11: Safety Tips

    • Keep the display dry if it is part of a surgical system. The display lacks protection against liquids or spills. • In locations where 240V outlets are used, connect the Dome Ex display only to a center-tapped, 240V, single-phase supply (for Canada and the United States). GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTABLE MARKED “HOSPITAL ONLY”...
  • Page 12 Unpacking and handling tips The Dome Ex display is a precision instrument that requires proper care to maintain product operation and adherence to specification. Unpack the display and components carefully, then set up and handle the unit properly to avoid damage to the LCD panel.
  • Page 13: Cleaning The Display

    Cleaning the display Observe the following guidelines to maintain the display and the LCD screen. • Use a clean, lint-free, absorbent cotton cloth to clear off any residual glue from removal of the protective film or to remove surface dust. Apply light pressure to remove the dust. •...
  • Page 14 Dome E2cHB Specification In locations where 240V outlets are used, connect the Dome E2cHB display to a center- tapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel Screen size, diagonal 498 mm (19.6 in.)
  • Page 15: Power Supply

    Power Supply WARNING: Use only the power adapter supplied with the Dome E2cHB display unit: the BridgePower BPM130S12F02 (130W) model. Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 3.0 A (3.0–1.5 A) Frequency 50–60 Hz...
  • Page 16 Dome E3 Specification In locations where 240V outlets are used, connect the Dome E3 display to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel Screen size, diagonal 528 mm (20.8 in.)
  • Page 17 Power Supply WARNING: Use only the power adapter supplied with the Dome E3 display unit: the BridgePower BPM130S12F02 (130W) model. Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 3.0 A (3.0–1.5 A) Frequency 50–60 Hz...
  • Page 18 Dome E3cHB Specification In locations where 240V outlets are used, connect the Dome E3cHB unit to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel Screen size, diagonal 541 mm (21.3 in.)
  • Page 19 Power Supply WARNING: Use only the power adapter supplied with the Dome E3cHB unit: the BridgePower BPM150S24F05 (150W) model. Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 2.5 A (2.5–1.5 A) Frequency 50–60 Hz ±...
  • Page 20 Dome E5 Specification In locations where 240V outlets are used, connect the Dome E5 display to a center-tapped, 240V, single-phase supply only (for Canada and the United States). Category Characteristic Item Specification Technology TFT AMLCD Dual Domain IPS, 3 subpixels per pixel Screen size, diagonal 541 mm (21.3 in.)
  • Page 21 Power Supply WARNING: Use only the power adapter supplied with the Dome E5 display unit: the BridgePower BPM130S12F02 (130W) model. Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 3.0 A (3.0–1.5 A) Frequency 50–60 Hz...
  • Page 22 Landscape Dome E5 2048 x 2560 Portrait 32-bit* 2560 x 2048 Landscape 32-bit palette is used for 256 and 1024 shades of gray. † 32-bit palette is used for 16.7 million and 1.06 billion colors. 20 | Dome Ex Display...
  • Page 23: Connector Ports

    Connector Ports All Dome Ex displays have a DVI connector. The Dome E3cHB unit also has a DisplayPort connector. Display Video Port DIN Connector Power Input Adapter Dome E2cHB, 8-pin 12V ±5% BPC BPM130S12F02 Dome E3, or (130 W) (with power switch)
  • Page 24: Led Status Lights

    LED Status Lights The two LEDs on the back panel of the Dome Ex displays provide information about the display. • LED A describes the digital-link status between the display controller and the interface. • LED B describes the display-panel status. it shows any faults currently in the panel.
  • Page 25 Flashing green Fault Solid yellow DDC power-on, 12V/24V power not detected Alternating green/yellow POST failure NOTE: A red LED or combination of red LEDs indicates a hardware or software fault. Contact Dome Technical Support for assistance. Technical Information | 23...
  • Page 26: Troubleshooting

    Windows Display Properties dialog. A single-link commodity When using commodity graphics graphics board is in use. boards with Dome E3, Dome E3cHB, and Dome E5 displays, be sure to install a DualLink DVI graphics board. 24 | Dome Ex Display...
  • Page 27: Regulatory Compliance

    (EMC) laws. WARNING! Users of Dome E3cHB and Dome E5 Displays The Dome E3cHB and Dome E5 displays are Class A products. In a domestic environment, either of these products may cause radio interference, in which case the user may be required to take adequate measures.
  • Page 28 Annex IX Classification Criteria rule 10, dash three. The above intended use for review, analysis, and diagnosis is a Class IIa medical device. Thus, we believe the Dome E3 is a Class IIa medical device, and subject to Notified Body’s approval.
  • Page 29 Furthermore, the applicable safety standards for an MDD Class IIa display are EN 60601-1:2006 and IEC 60601-1:2005. To help the medical device designer evaluate the suitability of this display, NDSsi has also conducted EMC testing to IEC 60601-1-2 as it can be applied. The display with its power supply alone does not represent a functional medical device.
  • Page 30 Declaration of Conformity document, contact NDS Surgical Imaging. The Dome E2cHB, Dome E3, Dome E3cHB, and Dome E5 digital flat-panel displays meet the essential health and safety requirements of, are in conformity with, and the CE marking has been applied according to the relevant EU Directives listed below, using the relevant section of the following EU standards and other normative documents;...
  • Page 31 U.S. FCC Compliance Statement Class B Digital Device: Dome E2cHB or Dome E3 Display This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
  • Page 32 Class A Digital Device: Dome E3cHB or Dome E5 Display This device complies with Part 15 of the FCC Rules. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
  • Page 33: Electromagnetic Emissions

    Model Dome E series needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS; Portable and mobile RF communications equipment can affect the Dome E series.
  • Page 34: Electromagnetic Immunity

    2. ELECTROMAGNETIC IMMUNITY Guidance and Manufacturer’s Declaration 211; Electromagnetic Immunity The model Dome E series is intended for use in the electromagnetic environment specified below. The customer or the user of the display should assure that it is used in such an environment.
  • Page 35 Guidance and Manufacturer’s declaration–Electromagnetic Immunity The model Dome E series is intended for use in the electromagnetic environment specified below. The customer or the user of the Dome E series should assure that it is used in such an environment. 60601-1:2005 COMPLIANCE –...
  • Page 36: Recommended Separation Distances

    Between Portable and Mobile RF Communications Equipment and the Dome E Series The model Dome E series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the display can help...
  • Page 37 Each Dome Ex display has one or more of the following certifications. China China Compulsory Certification regulating safety and EMC. GB4943-2011 GB9254-2008 GB17625.1-2012 Class A Dome E3cHB and Dome E5 Germany TÜV Rheinland GM Mark for medical products. Japan Voluntary Control Council for Interference by information technology equipment sold in Japan.
  • Page 38: Standard Limited Warranty

    † If, within 5 years of initial purchase, the maximum output of the Dome Ex/Dome Sx color display (excluding the Dome E4c) is determined by NDSsi to be less than 250 cd/m², NDSsi will repair or replace the display at its sole discretion. The display must have been run in native color mode at or below the CXtra default brightness setting as configured from the factory.
  • Page 39 EXCLUSIONS – WHAT IS NOT COVERED 1 Any product with a defaced, modified, or removed serial number. 2 Damage, deterioration or a malfunction resulting from accident, misuse, neglect, fire, water, lightning, or other acts of nature, unauthorized product modification, or failure to follow instructions supplied with the product.
  • Page 40 (exclusive of taxes, installation and shipping related fees). • NDSsi provides a complimentary 90-day Out of Box Failure (OBF) replacement warranty from the date of shipment from NDSsi. 38 | Dome Ex Display...
  • Page 41: Product Line

    CONTACTING NDSSI CUSTOMER CARE PRODUCT LINE Dome® TOLL FREE (866) 961-9340 LOCAL (503) 620-3787 E-MAIL medicalsupport@ndssi.com HOURS . – 5 . Pacific Time, Monday – Friday 1 Contact NDSsi Medical Support during the hours listed above. Have your NDSsi model number and serial number available. You may be required to provide proof of purchase as a condition for receiving warranty service.
  • Page 42 NDSsi reserves the right to suspend warranty coverage to any customer who is in default of payment until such payment has been made. 40 | Dome Ex Display...
  • Page 43 European Union. Power cord, CN 903-0497 Medical-grade power cord for use in China. DVI cable 903-0538 Video cable for use with any Dome Ex display. DisplayPort cable 903-1138 Video cable for use with the Dome E3cHB display. Quick reference 020-0915 Installation booklet for Dome display systems.
  • Page 44 NDS Surgical Imaging, LLC 5750 Hellyer Avenue San Jose CA 95138-1000 USA www.ndssi.com NDS Surgical Imaging BV Nijverheidscentrum 28 2761 JP Zevenhuizen (ZH) The Netherlands +31 180 63 43 56 Tel +31 180 63 21 91 Fax...