POWERING DOWN THE SYSTEM

IMPORTANT:
The ScanX is designed to be left on continuously during the active day. At the
end of the day, or whenever desired, power down the system simply by pressing
and holding the membrane READY switch (
for approximately two seconds, until the green LED above the READY switch
extinguishes.

MAINTENANCE

Maintenance Procedures
The ScanX is designed for many years of trouble-free operation. Maintenance as
described herein is minimal.
IMPORTANT:
Cleaning the ScanX
Turn off the ScanX disconnect the line cord from the Mains wall outlet and dis-
connect the computer connection cable from the ScanX before cleaning. Wipe
the outside surfaces with a soft paper towel dampened with a disinfectant solu-
tion or non-abrasive household cleaner. Be careful not to allow solvents TO RUN
OR DRIP into the ScanX. This could cause damage to the ScanX. Allow to air dry
before plugging in or turning back on.
ScanX Cleaning Sheets cannot be used on the ScanX
Note:
(P/N B7100) because of height limitations
Cleaning the Plate Transport
Over time, small debris and dust may accumulate in the plate transport mecha-
nism causing a loss in image quality and possible damage to the PSPs.
To ensure optimal performance of the ScanX
should be cleaned once per week using a new ScanX Cleaning Sheet each time.
Sample sheets are included with the ScanX and additional sheets can be
purchased from your dealer. Refer to ACCESSORIES page 27.
Disinfecting the ScanX
No disinfection is required for the ScanX.
Cleaning the Plate Guides
Wash the Plate Guides with a non-abrasive cleaner and water. Dry thoroughly.
IMPORTANT:
Always use standard infection control procedures when handling devices that contact
Note:
the patient.
Disinfecting the Plate Guides
There is no reason to routinely disinfect the Plate Guide unless contamination
is suspected. If a Plate Guide has touched a contaminated surface, it may be
immersed briefly in a cold sterilant according to sterilant manufacturers directions.
After disinfection, clean and dry the Plate Guide using the instructions above.
Never power down the system during a scanning
session.
Do not spray solvents or liquid directly on the scanner.
Plate Guides must not be put in an autoclave.
24
) on the Keypad Control Panel
®