3.3

Status Display

The display panel lights and the auditory alarms operate as follows to indicate the operational status of the
vaporizer:
Note: If the dial is at the q setting when an alarm occurs, the auditory alarm is activated and the light
indicating the alarm condition flashes. When the
is continuously illuminated.
3.3.1

Warm-Up

The amber WARM-UP light is continuously illuminated when the
vaporizer is in its warm-up mode. The dial cannot be turned
from q when the vaporizer is in its warm-up mode.
3.3.2

Operational

The green OPERATIONAL light is continuously illuminated
when the vaporizer is in its operational mode. The dial can be
turned from the q setting when the vaporizer is in the
operational mode.
3.3.3

Agent Level Display

The LCD agent level display consists of a series of bars which
indicate the nominal volume of agent in the vaporizer sump.
The
symbol denotes a 390 ml volume of agent and the
symbol denotes a 60 ml volume of agent in the sump.
As the sump fills, the first agent level indicating bar appears in the LCD volume display when there is
between 60 and 80 ml of agent in the sump. As the level increases more bars are displayed as illustrated on
Fig. 6 and as the level decreases less bars are displayed.
When the bars are either level with or below the
one full bottle. When there are no bars visible, as illustrated on Fig. 7, the Low Agent Level alarm is
activated as described in Section 3.3.4.
Note: The agent level display is calibrated to be most accurate when the vaporizer is in a perfectly level
position. Deviations from a level position affect the accuracy of the agent level display and may activate the
LOW AGENT level alarm.
D-Tec Vaporizer
O & M Manual Part No. 1107-0622-000
is pressed, the auditory alarm is muted and the light
<
, the sump can accommodate 240 ml of agent, which is

Fig. 5 Warm-Up

Fig. 6 Operational
Page 7
August 2001

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

D-tec vaporizer

Table of Contents