Utility Operation
4.3

Auto F.D.C.

This function is used to calibrate the film density.
Output film density must correspond to the density gradation curve value that is determined
by using the measured 24-step density data as reference.
When the film manufacturing (lot) No. has been changed or when the output density has
changed, use this function to calibrate the density in the system.
The operation procedure is as follows.
1 Start up the Utility. (Refer to section 4.2.1.)
2 Press the
[Fig. 4.5] Utility Initial Screen (1/3)
3 Select a tray targeted for calibration.
[Fig. 4.6] Tray Select Screen
4-4
button on the Utility initial screen (1/3).
Two or more trays can be
selected at the same time.
006-235-30 2004.03