Download Print this page
Jadak Bar Code Scanner Installation Instructions

Jadak Bar Code Scanner Installation Instructions

Fujifilm sonosite

Advertisement

Quick Links

Bar Code Scanner (USB) 

Installation Instructions

Follow these instructions to connect and program any of these USB bar code scanners:
JADAK flexpoint HS-1M on the X-Porte, Edge
Symbol LS2208 on the NanoMaxx, M-Turbo, or S Series ultrasound system
When connected, the bar code scanner is listed on the system's USB or USB Devices setup page.
See the Bar Code Scanner User Guide from FUJIFILM SonoSite for information on safety and cleaning.
Configuring the JADAK bar code scanner
To connect and program the Jadak bar code scanner
1 Attach the scanner cord to the bar code scanner and to a USB port on the ultrasound system.
See JADAK flexpoint HS-1
For location of USB ports, see the ultrasound system user guide or Help (X-Porte only).
2 Turn on the ultrasound system.
3 Scan the following two bar codes in the order listed:
USB Kybrd
STX Prefix
ETX Suffix
DOS CTRL
ASCII MODE
4 To ensure the scanner has been configured to work with your ultrasound system, FUJIFILM SonoSite
recommends opening a patient information form on your system and scanning a bar code. Your system is
successfully configured when you scan a bar code and the patient ID (and associated patient information, if
any) populates the patient information form.
Configuring the Symbol bar code scanner
To connect and program the Symbol bar code scanner
1 Attach the scanner cord to the bar code scanner and to a USB port on the ultrasound system.
See Symbol LS2208 Quick Start Guide included with the bar code scanner.
For location of USB ports, see the ultrasound system user guide.
,
M
Quick Start Guide included with the bar code scanner.
M-Turbo, S Series, or NanoMaxx ultrasound system

Advertisement

loading

Summary of Contents for Jadak Bar Code Scanner

  • Page 1: Installation Instructions

    Configuring the JADAK bar code scanner To connect and program the Jadak bar code scanner 1 Attach the scanner cord to the bar code scanner and to a USB port on the ultrasound system. See JADAK flexpoint HS-1 Quick Start Guide included with the bar code scanner.
  • Page 2 2 Turn on the ultrasound system. 3 Scan the following 9 bar codes in the order listed: For instructions to scan, see Symbol LS2208 Quick Start Guide. Note: If you have trouble scanning a bar code, try holding the scanner farther away from the bar code or at a different angle.
  • Page 3 4 Scan additional programming bar codes as desired. 5 To ensure the scanner has been configured to work with your ultrasound system, FUJIFILM SonoSite recommends opening a patient information form on your system and scanning a bar code. Your system is successfully configured when you scan a bar code and the patient ID (and associated patient information, if any) populates the patient information form.
  • Page 4 X-Porte, Edge, M-Turbo, NanoMaxx, S Series, SonoSite, and the SonoSite logo are registered or unregistered trademarks of FUJIFILM SonoSite, Inc. in various jurisdictions. All other trademarks are the property of their respective owners. Copyright 2013 FUJIFILM SonoSite, Inc. All rights reserved.

This manual is also suitable for: