Nds surgical imaging Zero Wire Duo User Manual page 33

High-definition wireless video system
Table of Contents

Advertisement

EMC Tables
portable and mobile RF communications equipment and the product
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the product can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the product as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power (W) of transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Recommended separation distances between
Separation distance, in meters according to frequency of transmitter
150 kHz to 80 MHz
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
0.23
0.73
2.3
7.3
23
26

Advertisement

Table of Contents
loading

Table of Contents