CO SureCell Cross Sensitivity Data
Gas Type
Hydrogen Sulfide
Sulfur Dioxide
Nitrogen Dioxide
Nitric Oxide
Chlorine
Hydrogen
Ethylene
Ammonia
O
Cross Sensitivity Data
2
Gas Type
Hydrogen
Methane
Nitrogen Dioxide
XT Series Personal Gas Detector
Concentration Applied
(ppm)
25
50
800
5
2
100
100
100
Concentration Applied
100%
100%
25 ppm
Reading
(ppm CO)
0
0.5
20
8
0
20
85
0
Reading
(%v/v O
)
2
-9%
0
0
19