[3] Applying Control Solution: Shake
the control solution bottle. Discard the
first drop of control solution and wipe
the bottle tip clean using a clean tissue
or paper towel. Dispense a second drop
onto a clean surface, such as an unused
resealable plastic bag.
Bring the tip of the test strip to the control solution sample.
The AgaMatrix Presto System will automatically detect if a test
is performed with AgaMatrix Control Solution.
When control solution is applied, the visual fill window on the
AgaMatrix Presto Test Strip turns blue.
Your meter will beep once indicating that you should remove
the tip of the AgaMatrix Presto Test Strip from the control
solution sample.
18
This is the visual
fi ll window. DO
NOT PUT CONTROL
SOLUTION HERE.
Apply control
solution here.
The entire
visual fi ll
window will
turn blue.
On the AgaMatrix Presto Meter display,
the droplet symbols appear on the test
strip symbol.
[4] Viewing Control Solution Test Result:
The result will appear on the display and, if
the volume is enabled, your meter will beep
once. The word "control" will also appear on
the display.
Compare the result of your control solution
test to the range printed on the AgaMatrix
Presto Test Strip vial label for the control
solution level (normal or high) that you
used. Your control solution result should fall
within this range. The control solution test
result is stored in memory.
To deactivate the meter, remove
the used AgaMatrix Presto Test
Strip. Dispose of used materials
properly. If your control solution
test results are out of the range,
repeat the test. If the results from
the control solution tests continue
to fall outside that range, do not use
the AgaMatrix Presto System to test
your blood glucose. Call customer
service.
Normal: 80-120 mg/dL
High: 220-320 mg/dL
FOR EXAMPLE ONLY.
Refer to your test
strip vial label for
corresponding range.