Hitachi L-2100 Instruction Manual page 72

Table of Contents

Advertisement

FLUSH MODE: PARAM.=1,
1
Enter "1" for column flushing under constant conditions, or "2" for column flushing
using a gradient.
[PARAM]
[NO]
PROG.(NO.9)=2
SET DELAY TIME:
<1.0-600.0>
Input the time period from the end of analysis to the
start of column flushing.
COLUMN FLUSH:
1.000
Input column flushing conditions (flow rate).
FLUSHING TIME:
<1.0-600.0>
Column flushing time can be input within 1.0 to 600.0
min.
Monitor screen
Monitor screen
When the time for column flushing arrives, the flushing is
executed according to No.9 program.
NOTICE:
The program for column flushing is fixed at No. 9.
Upon selecting column flushing, program No. 9 is
automatically set for flushing.
When column flushing is set, program No. 9 is not
usable as a gradient program. But program No. 9 is
usable with a link program.
The pump is stopped automatically when program
No. 9 ends.
4 - 28
[--]
TIME (min)
600.0
FLOW
TIME (min)
30.0
[--]
[--]
[--]