Purge Flow Rate - Hitachi L-2100 Instruction Manual

Table of Contents

Advertisement

4.3.3

Purge Flow Rate

Set a flow rate when executing purge.
Monitor screen
UTLTY
UTILITY CONDITION=1 UNIT=2 FLUSH=
1
EVENT 4 LINK= OTHERS=6
Select CONDITION.
CONDITION SET Z-ADJ=1 MOTOR-CTRL=2
2
PURGE=3, CALIBRATION=4
Select PURGE.
PURGE FLOW:
<0.001-2.499>
2.499
Enter a purge flow rate.
Monitor screen
Pressing the Purge key in PUMP ON status activates purging at the set flow rate. Pressing
the Purge key again during purge will reset the purge function. During purge, the upper
pressure limit is set at 2.9 MPa (30 kgf/cm
2
0 MPa (0 kgf/cm
).
FLOW
2
) and the lower pressure limit at
4 - 23
[--]