Cooling Water: Temperature Rise

Figure 3-4: 8200 Compressor Water Discharge Temperature Increase (°F)
3-6
The temperature of the cooling water as it leaves the compressor should
not exceed 100°F.
Use the graph in Figure 3-4 to determine the rise in cooling water
temperature as it passes through the compressor. This information is
provided for plant engineering personnel to determine cooling water
requirements.
Find the temperature rise:
1. Draw a vertical line upward from the horizontal axis of the
graph at the water flow rate determined from the previous
section, until it hits the graph curve.
The example shows dashed arrows pointing upward to the graph curve
from 0.50 gpm on the water flow rate axis.
2. At the point which the dashed arrows intersect the graph curve,
draw a straight line to the left to obtain the increase in output water
temperature.
The example shows a temperature increase of 20°F.