Ir Sensor - Vaisala HUMICAP HMT140 Quick Manual

Humidity and temperature transmitter
Hide thumbs Also See for HUMICAP HMT140:
Table of Contents

Advertisement

____________________________________________________________________ English
To reset the battery meter:
urn off the HMT140. Make sure the configuration Cable is disconnected
1.
T
from the sensor.
2. Press and hold the service button.
3. Turn the HMT140 on while the service button is engaged.
4. Wait four seconds.
5. Release the service button. The battery meter has been reset.

IR Sensor

Use the following procedure to trigger a data packet transmission using the
IR Sensor:
1. Turn on the HMT140 and wait for the display to turn on then off.
2. Place your hand near the display. The display will turn on again and the
beeper will beep.
3. To trigger a data packet transmission, ensure the Configuration Cable is
disconnected from the HMT140 (Sensor mode).
4. Place your hand over the display again and leave it for 5 seconds. The
Logger will beep twice to confirm transmission. It will attempt to connect
and transmit data to the host.
VAISALA _________________________________________________________________
13

Advertisement

Table of Contents
loading

Table of Contents