40-channel differential multiplexer with automatic cjc
Table of Contents

Advertisement

Figure 10
Connections for AC volts verification
CH1
H L
H L
H L
CH1
CH2
CH3
4. Set the Model 27xx for the 100mV range; make sure that REL is disabled.
5. Source 1kHz and 50kHz AC voltages for each of the ranges summarized in
Model 27xx readings fall within stated limits.
6. Press the OPEN key to open Channel 1.
Table 4
ACV reading limits
ACV range
Applied AC voltage
100mV
100.0000mV
1V
1.000000V
10V
10.00000V
100V
100.0000V
750V
300.000V*
* If the 5725A amplifier is not available, change the 300V @ 50kHz step to 220V @ 50kHz. Reading limits for
220V @ 50kHz = 219.36 to 220.64V.
H L
H L
CH2
CH1
H L
H L
H L
H L
H L
H L
H L
H L
H L
CH4
CH6
CH9
CH12
CH5
CH7
CH8
CH10
CH11
Model 7708
1kHz reading limits
(1 year, 18°C to 28°C)
99.910 to 100.090mV
0.99910 to 1.00090V
9.9910 to 10.0090V
99.910 to 100.090V
299.60 to 300.40V
Output
HI
Output
LO
H L
H L
CH13
CH14
Table
50kHz reading limits
(1 year, 18°C to 28°C)
99.830 to 100.170mV
0.99830 to 1.00170V
9.98300 to 10.0170V
99.830 to 100.170V
299.27 to 300.73V
Amplifier (Connect to Calibrator)
Calibrator (Output AC Voltage)
Shielded
Cable