that has a blood pressure cuff, arterial catheter, or
intravascular line.
SOLUTION:
• Check that calculations have been corrected
appropriately for the relevant variable. In general,
calculated saturation values are not as reliable as direct
laboratory hemoximeter measurements.
• If there is excessive light, cover the sensor with opaque
material.
• Circulation distal to the sensor site should be checked
routinely. The site must be inspected every 8 hours to
ensure adhesion, skin integrity, and correct optical
alignment. If skin integrity changes, move the sensor to
another site.
• Try to keep the patient still, or change the sensor site to
one with less motion.
• Observe all instructions, warnings, and cautions in this
manual and in the directions for use of the sensor.
PROBLEM: A valid SpO
disappeared.
CAUSE:
• A BP determination on the same limb is in progress.
SOLUTION:
• An alarm message codes will appear on the screen, and
the audible alarm will sound immediately.
PROBLEM: An SpO2 PLACEMENT error has occurred.
CAUSE:
• Weak or "noisy" signal.
SOLUTION:
Check the patient.
• If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; or replace the
sensor.
• Change sensor type.
signal was present but has
2
47