Using the Monitor

Manual Mode

To start a determination, press the START/STOP BP key (29).
A normal, uninterrupted Manual cycle takes about 40
seconds. The cuff pressure must drop below 5 mmHg
(neonate) or 15 mmHg (adult) before another determination
can be started. BP information will be displayed for a
duration of time set by the user. The factory default setting is
2 minutes (refer to "Retain Values" in the Using the Menu
System section for information on how to change this
setting). The user can set the display duration in 2, 5, 10, 15,
20, 30, 45, 60, 90, 120 minute increments. If another
determination is initiated during the user-set timeframe, the
display will change. This applies to Manual and Vitals (UK: All
Obs) modes. After power-off, the operating mode returns to
the default setting of Manual. The default setting of Manual
can be overridden to return to the previous user-selected
setting (Auto or Manual) by selecting Set BP (UK: BP Mode)
under the Service menu.
Note: The START/STOP BP key is an on-off switch; pressing it
will stop any BP determination (Manual, Auto, Stat, or Vitals)
that is in progress.

Auto Mode

Auto BP determinations are started by selecting the AUTO
BP key (27) or the Auto button under the Set BP button (UK:
BP Mode) in the Main menu.
Selecting the AUTO BP key (27) brings up the Set BP menu
(UK: BP Mode) and automatically starts an Auto BP
determination as long as the Monitor is in Manual BP mode.
If the Monitor is already in Auto BP mode, selection of the
AUTO BP key (27) brings up the Set BP menu (UK: BP
Mode) without starting a new determination until the preset
time interval has expired. Pressing the START/STOP BP key
during a series of Auto BP determinations will cancel the
determination in progress.