Adjusting the Footplate Angle
NOTE: For this procedure, refer to FIGURE 8.
1. Flip footplate up.
2. Screw the adjustment screw in or out until the desired angle is achieved.
3. Repeat STEPS 1 and 2 for the other footplate.
Adjusting the Tension of the Flip Up Footplate
NOTE: For this procedure, refer to FIGURE 9.
NOTE: The tension can be adjusted to increase or decrease the rotation effort of the flip up
footplates.
1. Loosen the mounting screw on the
front rigging angle hinge to decrease
the rotation effort.
NOTE: DO NOT remove the footplate
mounting screw.
2. Tighten the front rigging angle hinge
mounting screw to increase the rotation
effort.
3. Repeat STEPS 1 and 2 for the other
footplate.
Part No 1145733
Manual Center
Mount Footrest
(shown away from Extension Tube)
Adjustment
Screw
FIGURE 8 Adjusting the Footplate Angle
9
Formula™ CG Tilt Manual Center Mount Front Rigging
Footplate
Center Mount
Front Rigging
Mounting Screw
FIGURE 9 Adjusting the Tension of the Flip
Up Footplate
Footplate