Checking the Unit Before Usage

Before turning on the power, check the followings :
External appearance
Printer paper
Power cord
After turning on the power, check the followings :
External appearance
Switches and lamps
Date and Time
Measurement
16
• The monitor is not deformed due to falling or other impact.
• The monitor is not dirty or wet.
• The power cord is not damaged and there is no loose connections.
• The cuff or cuff cover is not deformed or damaged.
• The cuff cover is installed properly. (Page 26)
• The specified type of printer paper is loaded, and there is enough paper.
• Check if the power cord is completely connected.
• The power cord is firmly connected to the connector on the monitor.
• Make sure the power cord is not damaged (no core-wire exposure, breaks, etc.).
• The power cord is placed where it will not be tripped over.
• Make sure there is no smoke or bad smell.
• Press the switches and make sure they are functional.
• If a lamp is supposed to light when a switch is pressed, check to make sure it
lights.
• The date and time are set correctly. (Page 13)
• Take a measurement and make sure the result is near the usual blood pressure.
• The volume for audio voice is OK.
• There are no strange noises during measurement.
• There are no indication of IP leakage.