Technical Data - Accu-Chek Integra Complete Instructions For Use

Table of Contents

Advertisement

9.2 Troubleshooting
This message/
symbol appears ...
and means:
The meter and the infrared cable are too
far away from one another or not correctly
aligned.
Decrease the distance between the infrared
windows of the two devices to less than
10 cm.
Make sure the two infrared windows are
pointing towards one another. Re-start the
data download.
A computer software error occurred.
Read the Manual for the download software.
Electrostatic discharge
Turn the meter off and on again.
Meter is defective.
Please contact Customer Support (see Chap-
ter 19 for details).
94
10

Technical data

Meter type
Catalogue number/serial
number
Test principle
Measuring range
Sample size
Measuring time
System operating conditions
Meter storage conditions
Accu-Chek Integra
See type plate on the back of the
meter.
Determination of glucose in fresh
capillary blood by reflectance pho-
tometry. When using different speci-
men material please refer to the
package insert that came with the
Accu-Chek Integra test strip drums.
0.6–33.3 mmol/L (10–600 mg/dL)
Approx. 1.5 µL (1 microlitre =
1 thousandth of a millilitre)
Approx. 15 seconds (depending on
concentration)
+10 °C to +40 °C
Without batteries and test strip
drum: -40 °C to +70 °C
With batteries but without test strip
drum: -10 °C to +50 °C
With batteries and test strip drum:
+2 °C to +30 °C
95