F O R U S E The Game Ready System combines cold and compression therapies. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings.
INDICATIONS FOR USE _______________________________________ 1 CONTRAINDICATIONS ________________________________________ 4 GENERAL WARNINGS AND PRECAUTIONS _____________________ 4 THE GAME READY SYSTEM Your Game Ready System _____________________________ 7 Control Panel/Basic Controls ___________________________ 8 Priming a New Wrap __________________________________ 8 Control Unit Set Up ___________________________________ 9...
(such as blisters, increased redness, discoloration, or other noticeable skin changes), patients are advised to discontinue use and consult a physician. • The Game Ready Wrap is not sterile; do not place directly against open wounds, sores, rashes, infections, or stitches. May be applied over clothing or dressing.
• Always turn off the System and disconnect the power line cord from its electrical outlet when not in use or before adding ice and water. • Do not use any adapter other than that provided by Game Ready. Use of other adapters may result in electrical shock and will void the Game Ready warranty.
I N F O R M A T I O N 1. Protection against electric shock (5.1) The Game Ready power supply (Ault Model MW128) is considered to be Class I (grounded) equipment. The Game Ready power supply (GlobTek Model GTM21097-5012) is considered to be Class II (ungrounded) equipment.
YOUR GAME READY SYSTEM Control Unit Heat Exchanger Sleeve Carry Bag (optional) Connector Hose AC Adapter User’s Manual CONTROL UNIT ORIENTATIONS Figure A: Upright Figure B: Side The Control Unit can sit upright (Fig. A) or on one side (Fig. B) to fit on a shelf.
CONTROL PANEL / BASIC CONTROLS Pressure display Temperature display Green indicates unit has power Second tick marks Temperature knob Power (on/off) Celsius/Fahrenheit Timer on Alarm silencer Pressure setting Start/pause Decrease time Increase time Timer setting P R I M I N G N E W W R A P C O O L I N G...
C O N T R O L U N I T S E T Filling the Ice Box Stand unit upright. Then unplug the unit from the electrical outlet. To open the ice box lid, pull up on the latch on the end panel of the Control Unit. Fill the ice box with as much ice as possible.
• The Game Ready Wrap is not sterile; do not place directly against open wounds, sores, rashes, infections, or stitches. May be applied over clothing or dressing.
Air Pressure: When operating on low, medium or high pressure, the Game Ready Control Unit is designed to alert the user if it does not achieve its minimum or maximum pressure requirements.
G E T T I N G S T A R T E D 1. Pull up the latch to open the ice box lid. Fill the Control Unit with as much ice as possible and water, to 1” from the lip of the ice box. Close the lid securely. 2.
Page 14
CONTROL PANEL Pressure display Temperature display Green indicates unit has power Temperature knob Second tick marks Power (on/off) Celsius/Fahrenheit Timer on Alarm silencer Pressure setting Start/pause Decrease time Increase time Timer setting PRESSURE SETTINGS No pressure Medium High 5-15 mmHg 5-50 mmHg 5-75 mmHg PRE-SET PROGRAM DISPLAY...
A D J U S T I N G S Y S T E M S E T T I N G S ADJUSTING PRESSURE Press the pressure button on the control panel to adjust the level of compression. The system has four pressure options: 1. No pressure: circulates the ice water only, 2.
T H E B A G To operate the Game Ready System, you don’t need to remove it from the bag. Simply unzip the bag’s main compartment and end panel. To fill with ice and water, open the ice box lid. Attach the Connector Hose and the AC adapter to the end panel of the Control Unit.
- 122 Relative Humidity: 10% - 95% non-condensing. Store the Game Ready System in the carrying case or on a shelf, preferably with the lid open to allow the interior to dry. Caution: Do not keep in extreme hot or cold temperatures (above 120°F or below 33°F).
Bag Care and Storage Wipe clean with mild soap and water, air dry. Caution: Store the Game Ready System out of the reach of children and pets. HOW TO PACK THE GAME READY SYSTEM User’s Manual User’s Manual...
When you have compression. This often indicates determined if the problem is isolated that there is a leak in the to one Wrap or not, call Game Ready pneumatic compression circuit, customer service at 1.888.GameReady either in the Connector Hose or (426.3732).
Page 20
T R O U B L E S H O O T I N G ERROR SCREEN WHAT DOES IT MEAN? WHAT SHOULD I DO? • Make sure there is water in the ice Dry Pump Alarm: The Control Unit box.
Page 21
3 Check the ice box gasket to be sure it’s seated properly around the lip of the ice box. If it is not flat in the groove, push it down so that it is flush with the lip of the ice box. If the gasket is damaged, call Game Ready customer service at 1.888.GameReady for further assistance.
WARRANTY: CONTROL UNIT CoolSystems warrants that the Game Ready Control Unit, if properly used, will be free from defects in material and workmanship for a period of one (1) year after the date the Game Ready Control Unit was purchased.
Surge protection is recommended to protect the Control Unit. If the user of the Game Ready System requires continued operation during power mains interruptions, it is recommended that the Game Ready System be powered from a uninterruptible power supply. P R O D U C T...