Testing The Fuses; Replacing The Batteries And Fuses; Service And Parts - Fluke 15B User Manual

Hide thumbs Also See for 15B:
Table of Contents

Advertisement

3.
Soak a new swab with isopropyl alcohol and work
around the inside of each input terminal.
4.
Use a new swab to apply a light coat of fine machine
oil to the inside of each terminal.

Testing the Fuses

1.
Turn the rotary switch to
2.
Plug a test lead into the
probe to the A or mA  terminal.
A good A terminal fuse is indicated by a reading
between 000.0 Ω and 000.1 Ω. A good mA 
terminal fuse is indicated by a reading between
0.990 kΩ and 1.010 kΩ.
If the display reads , replace the fuse and test
again.
If the display shows any other value, have the Meter
serviced. See "Service and Parts" later in this manual.

Replacing the Batteries and Fuses

XW Warning
For safe operation and maintenance of the
product, remove batteries to prevent battery
leakage and damage to the Product if it is not
used for an extended period.
To prevent possible electrical shock, fire, or
personal injury:
.
terminal and touch the
Replace a blown fuse with exact
replacement only for continued
protection against arc flash.
Use only specified replacement fuses.
Batteries contain hazardous chemicals
that can cause burns or explode. If
exposure to chemicals occurs, clean with
water and get medical aid.
Replace the batteries when the low
battery indicator shows to prevent
incorrect measurements.

Service and Parts

If the Meter fails, first check the batteries and fuses, then
review this manual to make sure that you are operating
the Meter correctly.
To contact Fluke call:
+86-10-65123435 ext 15 in China
+91-11-450-94781/98200 29770 in India
+81-3-3434-0181 in Japan
+85-276-6196 in Singapore
+1-425-446-5500 anywhere in the world
Visit Fluke's Web site at
www.fluke.com
Multimeters
Service and Parts
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

17b

Table of Contents