Model 77 Series IV
Calibration Information
Performance Tests
When calibrating or measuring ac current, avoid the potential for errors from coupled noise by either:
1. Shorting the VeG to the COM input
or
2. Having no connection to the VeG input.
1. Set the Meter's Rotary Switch to the position called for in Table 2, and for all measurements other than
current (amps), connect the calibrator to the VeG and
If testing the milliamps function, connect the calibrator to the Meter
terminals.
If testing the amps function, connect the calibrator to the Meter
2. Referring to Table 2 or Table 3, apply the indicated calibrator output voltages to the UUT for each test.
3. Verify that the UUT displays a reading that is within the limits shown in the Meter Response column.
Steps
Test
(Switch Position)
e
1
[2]
Ohms
2
3
4
5
R
6
Continuity
7
K
8
AC Volts
9
10
11
12
13
14
15
16
K
17
Hz
AC Volts Frequency
12
Note
Table 2. Performance Tests
5500 Output
500 Ω
50 kΩ
5 MΩ
10 MΩ
40 MΩ
25 Ω
250 Ω
50 mV 45 Hz
5 V 45 Hz
5 V 1 kHz
50 V 45 Hz
50 V 1 kHz
500 V 45 Hz
500 V 1 kHz
1000 V 45 Hz
1000 V 1 kHz
5 V 50 kHz
[1]
input terminals.
COM
400 mA
and
10A
COM
Meter Response
Lower Limit
497.3 Ω
49.74 kΩ
4.974 MΩ
9.79 MΩ
39.19 MΩ
Beeper On
Beeper Off
0.47 V ac
4.898 V ac
4.898 V ac
48.98 V ac
48.98 V ac
489.8 V ac
489.8 V ac
978 V ac
978 V ac
49.94 kHz
and
input
COM
input terminals.
Upper Limit
502.7 Ω
50.26 kΩ
5.026 MΩ
10.21 MΩ
40.81 MΩ
0.53 V ac
5.102 V ac
5.102 V ac
51.02 V ac
51.02 V ac
510.2 V ac
510.2 V ac
1022 V ac
1022 V ac
50.06 kHz