DHP 1 Horsepower Owner's Manual page 7

Dental air compressor

Advertisement

DHP Air Compressor Maintenance - Section 7
The DHP Dental Air Compressors are all reduced maintenance products. They are specifically
engineered for minimal maintenance based on standard projected use. Excessive use of the
DHP Dental Air Compressor will require more scheduled maintenance than the standard recom-
mended maintenance listed below.
Daily Maintenance-Dental Air Compressor
1) The DHP Dental Air Compressor does not require any daily maintenance, however, it is strongly
recommended that periodic visual inspections as well as listening for any ' unusual ' noises be
done on a daily basis. The DHP Dental Air Compressor should operate (cycle on) when in use.
If the air compressor cycles on when dentist has not used any air out of the storage tank their may
be an air leak in the air system. If you notice any ' unusual ' noises or cycling problems contact
DHP customer service immediately.
Weekly Maintenance-Dental Air Compressor
1) The DHP Dental Air Compressor does not require any weekly maintenance, however it is
strongly recommended that periodic visual inspections as well as listening for any 'unusual '
noises be done on a weekly basis.
Monthly Maintenance-Dental Air Compressor
1) Visually inspect the oil level and add oil as required. See Diagram below
DO NOT OVER FILL THE OIL IT WILL CAUSE SERIOUS DAMAGE TO THE MOTOR
DO NOT USE OTHER INDUSTRIAL OIL IN THE DHP DENTAL AIR COMPRESSOR IT WILL
CAUSE SERIOUS DAMAGE TO THE MOTOR

Advertisement

loading

Related Products for DHP 1 Horsepower