Troubleshooting: No Power; Step 1.0 Circuit Breaker Reset; Step 1.1 Test Joystick; Step 1.2 Battery Test - Sunrise Medical Quickie S-11 Service Manual

Quickie s-11
Hide thumbs Also See for Quickie S-11:
Table of Contents

Advertisement

1.0 Circuit Breaker Reset
If On/off button is pressed and no light or bar is shown, check
for tripped circuit breaker (see figure A1.0.1) and make sure
all connections are clean and tight (including the batteries). If
the problem persists, then perform the following diagnostics.
1.1 Test Joystick
Check that the voltage is going to the controller, set the
meter to dc volts and take a voltage reading from pin 1
(using the red lead) and pin 2 (using the black lead) to the
charger port of the VSI controller (see figure A1.1.1) If the
voltage meter reads approximately 24 volts, replace the con-
troller, if the meter reads less than 12 volts, proceed to the
next step.
Note: Make sure the polarity is correct.
ing is intermittent, there is a connection or Controller
problem.If polarity is reversed proceed to step 1.6
If the total Battery Voltage reads approx. 24 V, and the Polarity is correct, replace the controller.
1.2 Battery Test
Check that the batteries are fully charged and in good condi-
tion. Remove the seat, and the battery cover, with controller
connected and turned on, use the meter to check the volt-
age across the battery terminals (see figure A1.2.1). If the
voltage meter reads between 12 -13.5 volts, then proceed to
next step. If the voltage meter reads below 12volts,charge
the batteries.
Note: To find a bad battery, use a battery load tester.
If the voltage meter reads below 12 volts, charge the batteries.
SUNRISE MEDICAL
Section 1

Troubleshooting: No Power

If the read-
QUICKIE S-11 SERVICE MANUAL
2
1
3
2005
A1.0.1
OFF
750
V DC
1000
V AC
200
200
20
200u
2000m
2000u
200m
20m
20M
200m
2000K
10A
200K
A DC
20K 2000
200
OHM
V mA
COM
10A
DIGITAL MULTIMETER
A1.1.1
A1.2.1