Item 707A, 707AX, 707AC Digital Blood Pressure Monitor with Manual Inflation & Automatic Deflation User Manual Read this manual before operating the Labtron Digital Blood Pressure Monitor. Save this manual for future reference. 707A-INS-LAB-RevD11...
SAFETY GUIDELINES - PLEASE READ BEFORE USE Important! Read and understand these instructions before using the Labtron Digital Blood Pressure Monitor. If you do not understand any part of these instructions, contact your medical professional or Graham-Field dealer for direction in the use of this product. WARNING: If components are damaged or missing, contact your Graham-Field dealer immediately.
FEATURES bulb Main features of cuff the Digital Blood Pressure Monitor, exhaust valve referred to in this button manual, are shown in the picture at right. cuff hose ON/OFF button Digital Blood Pressure Monitor Labtron Digital Blood Pressure Monitor...
BATTERY INSTALLATION / REPLACEMENT 1. Slide the battery cover off in the direction of the arrow. 2. Install or replace the four AAA batteries, oriented as shown by battery compartment diagram, into the battery compartment. 3. Replace the battery cover. Ensure that the battery cover is securely fastened.
in the arteries occurring when the heart contracts. The lower reading (diastolic pressure) shows the lowest pressure in the arteries, which occurs right before the heart contracts. Blood pressure readings are written with the highest value first, then the lowest value. Readings of 120/80 are considered to be normal, with high blood pressure being defined as a systolic pressure which is 140mmHg or more at rest and a diastolic pressure which is 90mmHg or more at rest.
Page 7
beverages, before a measurement. 4. Remove any clothing on the upper arm to place the cuff directly on the skin. Constriction of the upper arm by clothing or a rolled-up sleeve, as shown at excessive right, may cause an inaccurate reading; remove any constriction constricting garments.
Page 8
7. Attach the cuff: Hold the area of the cuff where the hose exits the cuff firmly against the inside of the bare upper arm directly over the brachial artery, and the bottom of the cuff approximately one inch (2-3cm) above the elbow, as shown in the third picture at right.
Measuring blood pressure Symbol key: LCD symbols you may see during measurement Deflating Inflating Measurement error Measuring Low battery 1. Ensure that cuff is properly placed and that you are positioned as described in previous paragraphs. 2. Press the ON/OFF button. All LCD symbols will appear momentarily.
Page 10
Info: The low ba ttery symbol will also appear at this time while the monitor checks the power level. This does not indicate that the batteries are low unless the low battery symbol remains on continuously. 3. After "0" displays, pressurize the cuff by squeezing the bulb until the pressure value displayed is 50mmHg above your usual systolic pressure.
Page 11
Info: To stop measurement in progress for whatever reason, press the exhaust valve button to release the pressure and turn off the device. Info: If during measurement the power drops to an insufficient level, the device will end the measurement and display the low battery symbol.
Page 12
7. If performing measurement for the first time, repeat the measurement two or more times to ensure that you and cuff are positioned correctly and that results are accurate and consistent. Record your results; there is a log provided, with examples, at the end of this manual for this purpose.
MAINTENANCE s NOTICE: To prevent product damage, please follow these recommended care and maintenance instructions: Do not drop or pull excessively on monitor components or subject them to strong impact. Do not use monitor close to any device with a strong electrical field such as television, microwave oven, X-Ray equipment, etc.
TROUBLESHOOTING Error indicators Indication Cuff pressure reaches 330mmHg Cause Twisted bladder inside cuff Correction Ensure that bladder lies flat within cuff Indication "Err" displays, cuff pressure releases, then "0 pressure" displays Cause Excessive arm movement or talking during measurement Correction Turn monitor OFF then ON and measure again while remaining still Indication Could not obtain pulse rate Cause...
General troubleshooting Problem Recommended Action Nothing appears Ensure that batteries are correctly oriented in battery holder (polarity in LCD matches battery case placement indicators) No measurement Ensure that cuff is positioned correctly occurs Go over measurement procedure; measure again Blood pressure Ensure that user and cuff are positioned as instructed measurements are Rest your arm on a table so that the cuff is level with your heart during...
707A, adult-size cuff, fits 7.5 in. - 12.2 in. (19-31cm) arm Monitor Model 707A X, large adult-size cuff, fits 11.8 in. - 17.7 in. (30-45cm) arm 707A C, small adult-size cuff, fits 5.1 in. - 7.9 in. (13-29cm) arm Measurement method...
Page 17
Specifications, continued * Environmental Operation 50°F ~ 104°F (10°C ~ 40°C), 10% ~ 85% RH conditions Storage & 14°F ~ 140°F (-10°C ~ 60°C), 10% ~ 95% RH transport Standard cited EN 1060-1 Noninvasive blood pressure measuring equipment general requirements EN 1060-3 Noninvasive blood pressure measuring equipment supplementary requirements for electromechanical...
WARRANTY GF Health Products, Inc. ("Graham-Field") warrants the Labtron Digital Blood Pressure Monitor Model 707A/707A X/707A C for a period of one year for defects in workmanship and materials. If a product is deemed to be under warranty, GF Health Products, Inc.