Specifications; Construction; Installation; Environmental Issues - Fluke 5686 User Manual

Glass capsule standard platinum resistance thermometer
Table of Contents

Advertisement

Specifications

Temperature range
Nominal R
Resistance ratio
Drift rate over the entire range
Drift rate over a range of 100 °C
Filling gas
Lead wires
Diameter (mm)
Length (mm)

Construction

Construction of the 5686 SPRT is shown in Figure 2. The 5686 covers the range from 13 K to
505 K. The 25.5 Ω sensor element is crafted using high purity platinum wire wound in a strain
free design on a fused silica cross frame.
Glass Sheath

Installation

Environmental Issues

Primary standard equipment should be used in a calibration laboratory or other facility
specifically designed for this purpose. Environmental requirements include:
• Stable temperature and humidity
• Clean, draft-free area
• Low noise level: low radio frequency, magnetic or electrical interference
• Low vibration levels

Mounting

Capsule-type SPRTs are usually used "totally immersed", meaning that they are generally inserted
into a well in a copper block. The connections are made to long fine copper wires insulated with
varnish or a similar coating, and these are thermally anchored to the block and to at least one
other point within the cryostat so as to reduce or eliminate heat flow into or from the thermometer.
For applications above 100 °C, the user should decide how to mount the capsule in his apparatus
according to the conditions.

Electrical Circuit

The 5686 SPRT is equipped with four platinum wires with Teflon sleeves. Four lead wires are
used to cancel lead wire resistance. For best results, the readout device should be equipped to
handle four-terminal resistors. The lead wire colors are red and black. Two red color leads are
from one end of the sensor resistance, and two black leads are from another end (See Figure 2).
Glass Capsule Standard Platinum Resistance Thermometer
Table 2. Specifications
–260 °C to 232 °C
25.5
tp
W(302.9146 K) ≥1.11807, W(234.3156 K) ≤0.844235
<0.005 °C per year
Typically 0.001 °C per year
Pure helium (99.999 %)
Four platinum wires, 3 cm long
5.5 ±0.3
56
Platinum Helix
Figure 2. The construction of 5686 SPRT
5686 Specifications
Ω
Sensor Support
Glass-Platinum Seal
Platinum Lead Wires
Specifications
5686.eps
5