Specifications And Model Selection Guide; Specifications; Cell Constants; Electrode Material - Honeywell 4909 CPVC Operation Manual

Conductivity cell removal/insertion assembly
Table of Contents

Advertisement

2. Specifications and Model Selection Guide

2.1 Specifications

Cell Constants

0.01, 0.1,1.0, 5, 10, 20, 25 and 50 cm

Electrode Material

Nickel or Platinum, as specified.

Wetted Parts

Cell Body:
Electrodes:
Mounting Materials:
Internal Sealing Materials: Viton, Teflon & EPDM

Pressure and Temperature (See Fig. 1-1)

125 psig (862 kPa) max. @ 23 F (-5 C)
90 psig (621 kPa) max. @ 122 F (50 C)
50 psig (345 kPa) max. @ 176 F (80 C)

Maximum Pressure

During Insertion or Removal: 50 psig.

Mounting

1-1/2" NPT male pipe. Overall length of removal device: approximately 20 or 22 inches (allow additional
clearance for cell withdrawal). See Figs. 3-1 and 3-2.

Purge Port

1/4" NPT female opening.

Insertion Depth

Varies between 4.5" and 6.8" nominal, depending upon cell constant. Greater insertion depths are optional.
See Figs. 3-1 and 3-2.

Electrical Connections

Three leads with integral automatic temp. compensator. Two leads without integral automatic temp.
compensator. Four leads with integral automatic temp. compensator)Table III = 333).
7/99