Error Messages

Code
Error Message
-109
Missing Parameter
-211
Trigger Ignored
-213
INIT Ignored
-224
Illegal Parameter Value
-310
System Error
+1500
External Trigger Source
Already Allocated
+2000
Invalid Card Number
+2001
Invalid Channel Number
+2006
Command Not Supported On
This Card
+2008
Scan List Not Initialized
+2009
Too Many Channels In Channel
List
+2011
Empty Channel List
+2012
Invalid Channel Range
+2600
Function Not Supported On
This Card
100 Matrix Modules Error Messages
Table C-2. Error Messages
Sending a command requiring a channel list without the channel list.
Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.
Attempting to execute an INIT command when a scan is already in
progress.
Attempting to execute a command with a parameter not applicable to
the command.
Too many characters in the channel list expression.
Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another switchbox.
Addressing a module (card) in a switchbox that is not part of the
switchbox.
Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g., channel 99 of matrix module).
Sending a command to a module (card) in a switchbox that is
unsupported by the module.
Executing an INIT without a channel list defined.
Attempting to address more channels than available in the switchbox.
No valid channels are specified in the <channel_list>.
Invalid channel(s) specified in SCAN <channel_list> command.
Attempting to begin scanning when no valid <channel_list> is defined.
Sending a command to a module (card) in a switchbox that is not
supported by the module or switchbox.
Potential Cause(s)