Timer/counter

Advertisement

Model 5304A
Maintenance
Table 9D-5-1. In-Cabinet Performance Checks (Continued)
Figure A
Figure H
HP 2??A
HP iBuA
CHAN A
V t R T
IN
PULSE
OUTPUT
oNC TEE
H P S 3 0 0 A
H P 5 3 0 4 A
TO C H A N A
soft
FEEDTHRU
b.
Set 222A for "+" PULSE POLARITY;
40 nsec pulse width; 150 mV p-p as
monitored on oscilloscope.
c.
Monitor 222A pulse output with oscil-
loscope and set REP RATE for 10 Hz.
d.
Set 5304A Channel A controls as follows:
ATT EN switch to XI
AC-DC to DC
SLOPE to +.
LEVEL to PSET or adjust until a
stable count is displayed.
e.
5300A display should be approximately
10 Hz.
f.
Repeat step c with 222A repetition rates
of 10 kHz, 100 kHz, 1 MHz and 10 MHz.
The 5300A display should be the same
as the 222A repetition rate setting.
g.
Set 222A PULSE POLARITY to
Set 5304A Channel A SLOPE to -.
Repeat steps b through f.
4. PER AVG A Mode.
RANGE: 10 Hz(.l sec) to 1 MHz (I /jsec).
Obtain the following test equipment:
HP Model 651B Test Oscillator.
HP
Model 11048B 50 OHM
Feed-thru
Termination.
a.
Connect equipment as shown in Figure B.
b.
Set 5304A Channel A controls as follows:
1. "Function" switch to PER AVG A.
2. ATTEN to XL
3. AC-DC to AC.
HP 6510
HP i BOA
CHAN
A
VERT IN
NPUT
BNC TEE
H P 5 3 0 0 A
H P 5304A
10MHz
FEEDTHRU
4. SLOPE to +.
5. LEVEL to PSET or adjusted until a
stable display is obtained.
c.
Monitor the oscilloscope and set 651B con-
trols for 10 Hz, 100 mV rms output.
d.
Display should be approximately * 00.0000
MS, C <* = OVERFLOW) (100 msec).
e.
Monitor the oscilloscope and set 651B for
the following frequencies; display should
be as listed.
65 IB
10 kHz
100 kHz
I MHz
5300A/5304A
100.000/isC
10.0000 f j s C
01.0000
M
s C
NOTE
Periods averaged are automati-
cally selected to provide maxi-
mum resolution.
5. T. I. A to B (Time Interval) Mode.
RANGE: 500 nsec to 10,000 seconds.
Obtain the following test equipment:
HP
Model
222A
Pulse
Generator
(2
required).
HP
Model
11048B 50-0hm
Feed-thru
Termination (2 required).
HP Model 180A Oscilloscope.
BNC Tee Connectors, (3 required).
a.
Connect equipment as shown in Figure C.
9D-5-17