:PULS:TRAN[1]:TRA:AUTO
:[SOURce]:PULSe:TRANsition[1]:TRAiling:AUTO
Set & Query
ON|OFF|ONCE
ON
Use this command to set/read the automatic coupling of the pulse trailing
edge transition time to the leading edge transition time.
ON
OFF
ONCE
To set leading and trailing edges independently:
:PULS:TRAN 6NS
:PULS:TRAN:TRA:AUTO OFF
:PULS:TRAN:TRA 15NS
:PULS:TRIG[1]:VOLT
[:SOURce]:PULSe:TRIGger[1]:VOLTage[:LEVel][:IMMediate]
[:AMPlitude]
Set & Query
TTL | ECL
TTL
Use this command to set/read the output levels at the TRIGGER OUT
connector.
The trailing edge transition time is automatically
set to the same value as the leading edge, and is
updated automatically each time the leading edge
transition time changes.
The trailing edge transition time is independently
programmable.
The trailing edge transition time is set ONCE to
the same value as the leading edge.
Set OUTPUT leading edge to 6 ns
Enable independent setting of trailing
edge
Set OUTPUT trailing edge to 15 ns