Using the Mixers with the 8566B Spectrum Analyzer

Set up the equipment

1. Connect the external harmonic mixer to the spectrum analyzer, as shown in
NOTE
Good-quality shielded SMA-type cables should be used to connect the mixer to
the spectrum analyzer to ensure that no signal attenuation occurs. Agilent
5061-5458 SMA-type cables may be used. Do not over-tighten the cables; the
maximum torque should not exceed 112 N-cm (10 in-lb.).
Figure 2-17
External Mixer Connection
CAUTION
Before connecting the Agilent 11975A Amplifier, set the ALC switch to ON.
Failure to do so can damage the mixer.
2. Power up the spectrum analyzer and then the amplifier.
3. Adjust the amplifier power to 16 dBm.
4. To access the spectrum analyzer's millimeter bands, press
until the desired band is displayed on the screen, as shown in
this is important because toggling through the bands, causes the band to automatically
harmonic lock as evidenced by the on-screen annotation of "Harmonic 6" changing to
Chapter 2
Using the Mixers with the 8566B Spectrum Analyzer
and then the up arrow
SHIFT
Figure
Operation
Figure
2-17.
2-18. Note that
45