Figure 2-10 Store and correct for conversion loss.
The second method for storing conversion-loss information lets you save individual
conversion-loss data points at specific intervals across the harmonic band, using
.
LOSS VS FREQ
To view or enter a conversion-loss data point:
a. Press
CNV LOSS VS FREQ
b. Enter the conversion-loss data at the frequency shown.
c. Use the step up
step.
Signal Identification
5. The IF output of a harmonic mixer contains many mixer products (frequencies of LO
source, 2LO
source, 3LO
harmonic band, a single input signal can produce many responses, only one of which is
valid.
These responses come in pairs, where the members of the valid pair are spaced 621.4
MHz apart (see
Figure
response (for this analyzer, the left member of a pair is not valid).
Chapter 2
Using the Mixers with the 856X Series Spectrum Analyzers
.
key to step through the band, entering the conversion loss at each
source...nLO
2-11) and the right-most member for the pair is the correct
source). As a result, within a single
Operation
CNV
33