en | Programming
4.2

First use

4.2.1

List of original factory settings

Parameter
* Locating mode
* Identification number
Assistance and fire priority
Special texts in German
Printer/display setup (Relay Plus Unit only)
Speaker volume
* Locating mode and identification number are set at first use. To change a value, reset the
unit. See Section 4.3.2 Resetting all the parameters and buffers, page 21.
4.2.2

Parameters set by the NurseCall Main Unit

The parameters listed below are programmed in the Main Unit and are applied to the Relay
Unit. See the NurseCall Main Unit user manual.
4.2.3

Locating mode and identification number

1.
2.
3.
4.
The identification number is attributed by the NurseCall Main Unit upon connecting the RS-
485 bus corresponds to the order of connection. It can also be set manually. The range for
this value is (A, B, C, ..., Y, Z, a, b, c, d, e, f) for a maximum of 32 Relay Units.
Automatic attribution by the NurseCall Main Unit
5.a. Press
5.b. Connect the NurseCall Relay Unit to the RS-485 bus. Press
enable the automatic attribution. The next free identification number is attributed by the Main
Unit and is displayed on the Relay Unit at the bottom right.
F.01U.252.722 | V1 | 2011.12
Display mode
Unit language
Date and time
Local acknowledgement and the access code
Output relay function and mode
Disconnect the NurseCall Relay Unit from the RS-485 bus.
Switch the NurseCall Relay Unit on with the switch located at the bottom of the unit.
Wait for the following display to appear:
LOCATING ?
( OFF / ON
)
Enable (ON) or disable (OFF) the indication of the transmitter position (locating mode)
on the NurseCall Relay Unit display. It must be the same as in the NurseCall Main Unit.
Press
to disable or
The identification number of the NurseCall Relay Unit must now be attributed.
RELAY UNIT ID: ·
(
+
/
-
)
or the Red button to erase any previous selection. The black box remains.
Original Factory Setting
On
No value
No
No
None
Midrange
to enable the locating mode.