  • Page 2 Introduction Thank you for choosing the Kogan NON-contact Digital Infrared Forehead Thermometer. With the unique technology, the forehead thermometer can give an accurate result via non-contact, without the interference from ambient temperatures. The product will self-test every time it starts up to ensure normal operation and accuracy.
  • Page 3: Table Of Contents

    Contents 1.The advantages of the Forehead Thermometer 1 2.Necessary safety instructions ......2 3.Instructions for product designs......4 4.How to measure the forehead temperature..4 5.How to measure the milk temperature....5 6.Instructions for display and operation....6 7.Technical specifications........9 8.After-sale service..........12 9.Security type.............12...
  • Page 4: The Advantages Of The Forehead Thermometer

    1.The advantages of the Forehead Thermometer Multiple uses are possible with the thermometer (measurement of body temperature and milk temperature) Forehead Thermometer provides measurement of body ℃ ℃ ℉ ℉ temperature, which is from 32 to 42.2 (from 89.6 to 107.9 ) , ℃...
  • Page 5: Necessary Safety Instructions

    Safe and hygienic ●No-contact measurement won’t bring about the accumulation of bacteria, so it doesn’t need to be clean on purpose practically. ●There is no danger of breaking glass. ●It is totally safe for children to use. Fever warning ℃ ℉...
  • Page 6 ●Please do not use the product if the temperature sensor or the forehead thermometer show any sign of damage. Do not try to repair the product if it is damaged. Please contact Kogan Customer Service team if you have any issues.
  • Page 7: Instructions For Product Designs

    3.Instructions for product designs (1)Temperature sensor (2) Press button for measuring body temperature (3) Press button for measuring milk temperature (4) LCD display screen (5) Battery cover 4.How to measure the forehead temperature The thermometer measures the infrared energy emitted from forehead, which will focus through the lens and convert to temperature values via the thermopile and measuring circuit.
  • Page 8: How To Measure The Milk Temperature

    5.How to measure the milk temperature The thermometer measures the infrared energy emitted from the feeding bottle, and it shall display the milk temperature by transition calculation and emission rate compensation. Please use the press button (3) for measuring. In order to avoid inaccuracy: ●Please make sure that there is no dirt in on the temperature sensor ●Please make sure that there is no intense emotion and movement...
  • Page 9: Instructions For Display And Operation

    6.Instructions for display and operation Operational method and Sound and backlit instruction for displays 1.Measurement of body temperature Forehead 1.Aim the temperature sensor temperature mode (1) at the patient' forehead 1.When directly (effective distance from temperature value 1 to 6 cm) under shutdown ℃...
  • Page 10 There shall be signal for relative mode under measurement ℃ ℉ preparedness. The signal shall twinkle. If the measurer presses button(2) or (3) again, the measurement mode is available. 1.The measured value exceeds ℃ ℉ /212.0 under milk 3 short tick temperature mode.
  • Page 11 Inquiry for memory data, Storing 20 groups of data Operational method and instruction for Sound and LCD Display displays backlit Click button (3) for 4-8 seconds and LCD displays “- - -” with M signal silence twinkling. Click button (3) again and the LCD displays the first data group with M signal twinkling.
  • Page 12: Technical Specifications

    Attention: ●Electromagnetic interference: The thermometer contains sensitive electronic component and should not be used under conditions with electromagnetic interference, (such as near mobile phones and microwaves.) ●Please dispose the used products and batteries in accordance with local regulation requirements when the products and batteries are not available.
  • Page 13 ℃ LCD displays “L ” if the measured temperature is below the minimum of measurement range. ℃ LCD displays “H ” if the measured temperature is below the minimum of measurement range. 7.Abnormal LCD displays “Er1”if the temperature measurement state display circuit is abnormal (fault of SENSOR or the temperature measurement circuit), or the ambient ℃...
  • Page 14 12.Current <2μA; I <0.5mA; I <2mA; stand-by working Buzzer on consumption <15mA banklight ≧ Red 1.2cd/m 13.LED backlit specifications Green≧2.0cd/m 15.Operational ℃ ℃ ℉ ℉ ASTM 10 -104 )/15-95%RH conditions 16.Type of Applicable for forehead temperature and milk measuing temperature 17.battery Changeable for two 1.5V triple A batteries 18.Battery life...
  • Page 15: After-Sale Service

    8.After-sale service Application for repair or replacement should be presented during the warranty period. The damage caused by improper use is not under warranty scope. Batteries and packaging are not under warranty scope as well. 9.Security type The signal indicates that the thermometer is a BF type device with internal power supply.