Ac Voltage Specifications - Fluke 8808A Datasheet

Extended specifications
Hide thumbs Also See for 8808A:
Table of Contents

Advertisement

AC Voltage Specifications

AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50kHz to 100 kHz, add 0.13 % of range.
Maximum Input ........................................................750 V rms or 1000 V peak or 8 x 10
Measurement Method .............................................AC-coupled true-rms. Measures the ac component of input with up to
AC Filter Bandwidth:
Slow .......................................................................20 Hz – 100 kHz
Common Mode Rejection .......................................60 dB at 50 Hz or 60 Hz (1 k unbalance)
Maximum Crest Factor ..........................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) .........Crest Factor 1-2, 0.05 % of full scale
Full-Scale
Range
(5-1/2 Digits)
200 mV
199.999 mV
2 V
1.99999 V
20 V
19.9999 V
200 V
199.999 V
750 V
750.00 V
Range
Frequency
200 mV
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
2 V
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
20 V
20 Hz – 45 Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
200 V
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
750 V
20 Hz – 45Hz
45 Hz – 20 kHz
20 kHz – 50 kHz
50 kHz – 100 kHz
Notes:
[1]
Accuracy given as ± (% of reading + % of range)
4 Fluke Corporation
Fluke 8808A Digital Multimeter Extended Specifications
1000 V dc bias on any range.
Crest Factor 2-3, 0.2 % of full scale
Input Characteristics
Resolution
Slow
Medium
1 uV
10 uV
10 uV
100 uV
100 uV
1000 uV
1 mV
10 mV
10 mV
100 mV
Accuracy
Accuracy
90 days
23 °C ± 5 °C
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
0.8 + 0.05
0.15 + 0.05
0.3 + 0.05
0.8 + 0.05
7
Volts-Hertz product
Input Impedance
Fast
1 M ±2 % shunted by
10 uV
100 uV
1000 uV
10 mV
100 mV
[1]
1 year
Outside 18 – 28 °C
23 °C ± 5 °C
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
0.9 + 0.05
0.2 + 0.05
0.35 + 0.05
0.9 + 0.05
<100 pf
Temperature
Coefficient/°C
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
0.01 + 0.005
0.01 + 0.005
0.01 + 0.005
0.05 + 0.01
8808A Fluke Corporation 3