8808A Specifications
Voltage
100V Setting .............................................................90 V to 110 V
120V Setting ............................................................108 V to 132 V
220V Setting ............................................................198 V to 242 V
240V Setting ............................................................216 V to 264 V
Frequency ..................................................................47 Hz to 440 Hz.
Power Consumption ................................................15 VA peak (10 W average)
Dimensions
Height .........................................................................88 mm (3.46 in)
Width ..........................................................................217 mm (8.56 in)
Depth .........................................................................297 mm (11.7 in)
Weight ........................................................................2.1 kg (4.6 Ibs)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating................................................................0 C to 50 C
Storage....................................................................-40 C to 70 C
Warm Up.................................................................½ hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating................................................................Uncontrolled (< 10°C)
Storage ...................................................................-40 C to 70 C <95 %
Altitude
Operating ...............................................................2,000 Meters
Storage .......................................................................12,000 Meters
Vibration ....................................................................Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V.
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay ...........................................................400 ms
External Trigger Delay ..........................................<2 ms
External Trigger Jitter ...........................................<1 ms
Trigger Input ...........................................................TTL Levels
Trigger Output ..........................................................5 V max

Math Functions

Min/max, relative, hold, compare and dB functions.
Electrical
Input Protection ......................................................1000 V all ranges
Overrange ..................................................................10 % on the largest ranges of all functions except continuity and diode

Remote Interfaces

RS-232C

Warranty

One year
2 Fluke Corporation
Fluke 8808A Digital Multimeter Extended Specifications
<90 % (10 °C to 30 °C)
<75 % (30 °C to 40 °C)
<45 % (40 °C to 50 °C)
test