TEMPERATURE RECORDER (OPTION)
<MTR-G85>
1. Attach the recorder to the mounting kit (MTR-S7) by
following the procedure shown in the manual enclosed
with the mounting kit.
2. Remove the screw on the cover for the recorder space
on the upper front of the freezer. Open the cover. Fig. 7
3. Join the wire in the unit compartment with the recorder
wire by each wire connector.
4. The sensor of the temperature recorder is put in the
chamber through the access port in the top of the
product.
5. A recorder sensor is to do fix with a supplied nylon
clip the inside wall.
6. A mounting kit and a temperature recorder are
installed on the freezer.
7. Operate the freezer until the chamber temperature
gets to the set temperature.
temperature and chamber temperature displayed on
the control panel. Adjust the zero adjustment volume
on the temperature recorder so that the recorded
temperature can corresponds with the displayed
temperature if they are not compliance each other
See Fig. 8
See fig. 5.
See fig. 9
Check the recorded
Cover
27
Shaft
Fig. 7
Screw
Connector for
MTR-G85
Fig. 8
Fig. 9