Mitsubishi Electric P95DW Operation Manual page 8

Digital monochrome printer
Hide thumbs Also See for P95DW:
Table of Contents

Advertisement

Guidance and manufacturer's declaration - electromagnetic emissions
The Model P95DW/P95DE is intended for use in the electromagnetic environment specified below.
The customer or user of the Model P95DW/P95DE should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11/EN 55011
RF emissions
CISPR 11/EN 55011
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The Model P95DW/P95DE is intended for use in the electromagnetic environment specified below.
The customer or user of the Model P95DW/P95DE should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC/EN 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
Complies
IEC/EN 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input
±1 kV for input
/output lines
/output lines
±1 kV line(s) to
±1 kV line(s) to
line(s)
line(s)
±2 kV line(s) to
±2 kV line(s) to
earth
earth
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 0.5 cycle
for 0.5 cycle
40% U
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Electromagnetic environment - guidance
The Model P95DW/P95DE uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause interference in nearby
electronic equipment.
The Model P95DW/P95DE is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
)
T
environment. If the user of the Model
P95DW/P95DE requires continued
T
operation during power mains
)
interruptions, it is recommended that the
T
Model P95DW/P95DE be powered from
an uninterruptible power supply or a
T
battery.
)
T
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.