Important Note - Sunbeam 7652 Instruction Manual

Automatic digital blood pressure monitor

Advertisement

1. Turn Power ON
When power is turned ON, all display marks appear for approximately 1 second.
l
When the "0" is displayed the unit is ready for measurement. If the '7" mark appears, release air
l
in the arm cuff by squeezing it or removing the hose plug.
2. Attach the arm cuff to the left arm for best results.
3. Set the pressure value approximately 40 mmHg above your normal
Systolic blood pressure by pressing the PRESSURE SWITCH. There are 9
settings: 120,140,160,160,200,220,240,260,
Note" on following page.
4. Press the START SWITCH.
The arm cuff pressurizes when the Start Switch is pressed.
l
When the Start Switch is subsequenHy pressed during pressurization, air in the arm
l
cuff is released.
5. When Pressurization is completed, the automatic exhaust mechanism will
gradually reduce arm cuff pressure, and the "v" mark will appear indicating that
measurement is in progress.
1) Pressure value appears on the left, and exhaust speed on the right side of the display.
Current Pressure Value
(gradually
being
reduced)
2) When a pulse is detected, the pressure value moves to the right side of the display. At
this time the ' I e " mark flashes at the same rate as pulse rate, and the buzzer
sounds when th6 pulsation is sufficient.
Flashing

IMPORTANT NOTE

If insufficient pressure is obtained, the unit will inflate the cuff again to the next
l
pre-set pressure. The unit will continue to reinflate until it is able to obtain an
accurate pressure, each time inflating to the next preset pressure. At that point, if the
unit is unable to obtain sufficient pressure the "Err" mark will appear in the display.
6. Turn Power OFF
Press the Power switch to turn power OFF. (See Diagram A)
7. Previous Measurement Display
When the Memon/ switch is pressed, the previous measurement value is shown for
about 5 seconds.-
8. Automatic Power OFF Function
If this unit is left on after measurement an Automatic Power-OFF function turns the
power OFF after approximately 2 min. 30 seconds. To continue measurement press,
the power switch turning the unit ON.