Page 2
General Description This device measures blood pressure and pulse rate and saves the results in Memory. With proper use, the readings taken by this blood pressure monitor are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method.
SAFETY INFORMATION WARNING This device is intended for adult use in homes only. • This device is not suitable for use on neonatal patients, pregnant women, patients with The signs below might be in the user manual, labeling or other components. •...
Page 4
WARNING WARNING Please check that operation of the device does not result in prolonged impairment of patient It is recommended that the performance should be checked every two years and after • • blood circulation. maintenance and repair, by retesting at least the requirements in limits of the error of the The device cannot be used with HF surgical equipment at the same time.
The low pressure measured. Measurement Unit of the blood mmHg mmHg BLOOD PRESSURE MONITOR CUFF (TYPE BF APPLIED PART) pressure. (TMB-1583-S) 8.7” - 16.5” (22CM - 42CM) Batteries are low and need to Low battery be replaced. Pulse/min Beats/minute Measurement Unit of Pulse Rate.
INSTALLING AND REPLACING THE BATTERIES LCD DISPLAY SIGNAL LCD DISPLAY START/STOP BUTTON SET BUTTON LOCK BUTTON MEM BUTTON CUFF TUBE • Open the battery cover. • Install the batteries by matching the correct polarity, as shown. • Replace the battery cover. REPLACE THE BATTERIES WHENEVER THE FOLLOWING HAPPENS: •...
SETTING DATE AND TIME SELECT THE USER ID It is important to set the Date and Time before using your blood pressure monitor, so that a time stamp can be assigned to each record stored in the memory. 1. When the monitor is off, press and hold “MEM”. The user ID will show. (The setting range of the year: 2021—2061, Time format: 12/24H) Press ”MEM”...
ATTACHING THE CUFF TAKING THE MEASUREMENT 1. Remove all jewelry, such as watches and bracelets, from your left arm. There are two ways to take a measurement: Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right 1.
RECALL THE RECORDS DELETING RECORDS To delete all results for the selected user (User 1, for example.) 1. Press and hold the “MEM” button When the monitor is in memory recall mode, the display will show ”dEL ALL” + User ID. 1.
KEY LOCK BUTTON TIPS FOR MEASUREMENT Measurements may be inaccurate if taken in the following circumstances: To help conserve batteries and electricity, you may lock the keys between use. This prevents the monitor from turning on if any key is pressed unintentionally. •...
WHAT IS SYSTOLIC PRESSURE AND DIASTOLIC IRREGULAR HEARTBEAT INDICATOR PRESSURE? An irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuring the systolic and diastolic blood pressure. During each measurement, the monitor records all of the Systolic Diastolic pulse intervals and calculates the average;...
Page 12
WHY DOES MY BLOOD PRESSURE FLUCTUATE PROBLEM DISPLAY CAUSE SOLUTION THROUGHOUT THE DAY? Batteries are Replace with new exhausted. batteries. 1. Individual blood pressure varies naturally throughout the day. 2. Cuff placement and body position can also affect readings, so try to take the measurement Batteries are inserted Insert the batteries Display will not...
LIMITED WARRANTY Battery powered mode: DC 6V (4 × 1.5V AAA batteries) Power supply AC adapter powered mode: 6V 1A (Please only use the recommended AC adapter). This product has a limited warranty of 3 year(s) on the blood pressure monitor, Display mode Digital LCD Display V.A.
AUTHORIZED COMPONENT COMPLIED STANDARDS LIST Please use the authorized adapter. EN ISO 14971:2019 / ISO 14971:2019 Medical devices - Application of risk Risk Management management to medical devices EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices. Symbols to be used Labeling with medical device labels, labelling and information to be supplied.
Voltage dips, short interruptions 135°, 180°, 225°, 270° and 315°. 135°, 180°, 225°, 270° and 315°. the equipment TMB-1583-S including cables specified by the manufacturer. Otherwise, and voltage variations on power 0% UT; 1 cycle and 70% UT; 0% UT; 1 cycle and 70% UT;...
Page 16
Table 3 Guidance and manufacturer’s declaration - electromagnetic Immunity Radiated RF Test IMMUNITY Band Modula- Distance IEC61000-4-3 Frequency Service Modulation TEST LEVEL (MHz) tion (W) (Test (MHz) (V/m) specifications Pulse ENCLOSURE 380-390 TETRA 400 modulation PORT b) 18 Hz IMMUNITY to RF wireless communicati- FM c) ±...