Cautions; Procedures For Temperature Sensor Calibration; Remote Warning Terminals - Haier HYC-260-360 Operation Manual

Pharmaceutical refrigerator
Table of Contents

Advertisement

1) The appliance should be connected to a dedicated socket for 220V/50Hz power
supply.Allow some time lapse before loading the appliance.
2) The warning sound signal may be heard when the appliance is first connected to the
power source and can be inactivated by pressing the buzzer buttom. The warning
indicator stays on until the temperature of the test bottle sensor reaches 5
3) Ensure that the two test bottles are filled with 10% glycol solution.(see article 4,section
installation on page 6).
4) The appliance is preset during manufacture for operation temperature 2~8 .